The BT units are constant temperature water baths that shake.
They feature an easy-to-use and flexible inner tank and integrated vibrating design. They also incorporate a thermostat that enables fixed operation and comprehensive safety features.
Model
100
200
300
Temp. Control Accuracy
Room Temp. +5deg.C to +80deg.C
Temp. Distribution Accuracy
±0.1
Accommodation Capacity (Test tube, dia.16.5mm)
130 pcs.
169 pcs.
260 pcs.
·The vibration range varies extends from 10 to 44mm and can be freely changed, while the oscillation frequency ranges from 20 to 160 times per minute,allowing for non-incremental speed transition. settings are input using a controller with a digital display.
·Settings for the temperature and timer can be easily input using the up and down keys, and the settings appear on a digital display.
Specifications
Model
BT100
BT200
BT300
Performance
Method
Reciprocating shaking system, Stirring by pump
Operating temp. range
Room temp. +5deg.C to 80deg.C
Temperature control accuracy
±0.02deg.C to 0.08deg.C
Temperature distribution accuracy
±0.1deg.C
Temperature rising time
Approx. 95 min.
Approx. 70 min.
Approx. 120 min.
Shaking frequency
20 to 160 times/min.
Shaking width
10 to 40mm (adjustable)
Components
Temp. Controller
PID control by microprocessor
Temp. Sensor
Pt resistance thermometer (double sensor)
Temp setting method
Digital setting
Temp display method
Digital display
Overheat protector
ON/OFF control by microprocessor
Overheat protect setting method
Digital setting
Overheat protect sensor
K-thermocouple (double sensor)
Heater
Copper pipe heater (nickel plated)
1.2kW
1.9kW
2.5kW
Stirring System
Magnet pump
6W
10W
10W
Shaker
Gear motor
25W
40W
40W
Shaking speed setting system
Analog setting
Shaking speed display system
Digital display
Timer
1 min. to 999 hr. 50 min.
Digital display, Quick automatic start/stop
Safe Device
Safety countermeasures
Self-diagnosis function(Heater defective, Sensor defective, Set value abnormal, SSR short circuit and Overheat protector) Warning buzzer and alarm lump, Over current, short circuit breaker, Heater no-load operation prevention device and overheat protector.